Resolution between Domperidone and Droperidol

Home Resolution between Domperidone and Droperidol

Column : C18 100mm*4.6mm, 3µm
Flow Rate: 1.5ml/min
Wave Length : 280nm
Temperature: 25°C
Runtime : 20 minutes

Get In Touch

    Preparation of Mobile Phase :

    1.) Mobile Phase : A
         Dissolve 5.0g of Ammonium Acetate in 1000ml of HPLC Grade Water

    2.) Mobile Phase : B
          Methanol HPLC Grade

    3.) Gradient Program :

    Time (min)Mobile Phase-A(%V/V)Mobile Phase- B(%V/V)
    0-107030
    10-150100
    15-15.017030
    15.01-207030


    4.) Diluent :
    N,N Dimethylformamide
    Domperidone Reference Solution (a) : (0.10 and 0.15 mg/ml) Weight accurately about 10mg of Domperidone  CRS and 15mg of  Droperidol CRS into a 100ml Volumetric flask dissolve and dilute to volume with Diluent and mix Inject 10µl of Domperidone Reference Solution (a) in duplicate and calculate Resolution between Domperidone and Droperidol.

    Acceptance Criteria:

    Resolution between Domperidone and Dropridol should not be less than 2.0

    Column Name : Zodiac C18 Column

    Compound Name : Droperidol, Domperidone

    Get Your Quote or Call: 040-29881474

    We focus on supporting laboratory workflows & optimizing lab-wide operations