Column : Zodiac-Phenyl Column
Column Size : 250mmx4.6mm, 4µm
Mobile Phase :
M.P.: 0.77g/L of Ammonium acetate in water & adjust with formic acid to a pH of 5.60-5.65
Flow Rate : 10ml/min
Buffer Details : To separate related substances of Fluorouracil using a Zodiac Phenyl column in high-performance liquid chromatography (HPLC) as outlined in the USP method. The USP method for Fluorouracil typically employs a mobile phase that might include a combination of water, acetonitrile, and a buffer. Prepare a standard solution of Fluorouracil with a known concentration. This solution will be used to create a calibration curve and to identify peaks. Equilibrate the Zodiac Phenyl column with the mobile phase before use. Ensure the system is stabilized.
Detector : UV 254 nm
Column Temperature : 25°C
Sample Temperature : 25°C
Injection Volume : 10 μι
Diluent/Sample Solvent : Water