Related substances for Finerenone API by HPLC

Application ID : AN-1225/2

Analytes

  1. 1.Peak1 2.Finerenone 3. Peak2 4. Peak3
Conditions

Column : Presto 120 C8 AQ Column

Column Size : 150mm*4.6mm, 2.7µm

Mobile Phase :
Preparation of mobile phase-A : T Take 1000ml of water and
add1 ml of triflouroacetic acid and sonicare for degas.
Preparation of mobile phase-B : Mix Water: acetonitrile (50:50%
v/v) and sonicate for degas

Flow Rate : 0.5 mL/min

Detector : UV 242 nm

Elution Mode : Gradient program

Run Time : 55 min

Column Temperature : 20°C

Injection Volume : 10 µL

Description
Finerenone (often handled in its base form or specific salt configurations
like oxalate during chemical development) belongs to the drug category of
Non-Steroidal Mineralocorticoid Receptor Antagonists (MRAs).In accurate
analysis of the active pharmaceutical ingredient requires effective
separation of Finerenone from its related impurities.This HPLC application
demonstrates the separation of Finerenone using a Presto 120 C8 AQ
column. The method provides a clear chromatographic profile with
Finerenone eluting at approximately 11.318 minutes, while the observed
impurities are separated at approximately 10.026, 12.982 and 22.850
minutes.