Separation of Related Substances of Fluorouracil_USP Method

Application ID : 70

Analytes

  1. 1)Unknown Imp.1
  2. 2)Fluorouracil Related Compound A
  3. 3)Unknown Imp.2
  4. 4)Fluorouracil Related Compound B
  5. 5)Uracil
  6. 6)Fluorouracil
  7. 7)Unknown Imp.3
  8. 8)Fluorouracil Related Compound E
Conditions

Column : Zodiac-Phenyl Column

Column Size : 250mmx4.6mm, 4µm

Mobile Phase :
M.P.: 0.77g/L of Ammonium acetate in water & adjust with formic acid to a pH of 5.60-5.65

Flow Rate : 10ml/min

Description
To separate related substances of Fluorouracil using a Zodiac Phenyl column in high-performance liquid chromatography (HPLC) as outlined in the USP method. The USP method for Fluorouracil typically employs a mobile phase that might include a combination of water, acetonitrile, and a buffer. Prepare a standard solution of Fluorouracil with a known concentration. This solution will be used to create a calibration curve and to identify peaks. Equilibrate the Zodiac Phenyl column with the mobile phase before use. Ensure the system is stabilized.